# Calibration of Glove Tracker (Position Offset)


Using the functionality of DivingStation, you can calibrate the position of the Glove Tracker relative to the Glove by adjusting the position offset.

  1. From the left tab of the DivingStation screen, click on the Calibration button.
    hand_offset_1
  2. Once the calibration screen is displayed, select the controller from the options at the top.
    hand_offset_2
  3. When the screen switches as shown in the image, select the hand you want to offset using the checkbox and press the Start button.
  4. The position of the selected Controller (avatar's hand) in VR will be fixed, allowing you to move the Glove Tracker.

    TIP

    If both hands are selected, calibration will start from the left hand in sequence.

  5. Align the position of the Glove Tracker to the position of the Controller on VR.
    hand_offset_3
  6. Follow the on-screen instructions and confirm the position by gripping/clicking the button on the opposite hand.
  7. For both hands, calibration of the opposite hand's position will start next; please follow the on-screen instructions.
  8. Follow the on-screen instructions, grip one hand, and complete the calibration.

TIP

You can reset the position by pressing the Calibration > Controller > Reset button. [Refer to steps 2/3]